c. Exit the clock setting menu directly by press the "menu button". If you have reset the time or date,
when exiting the clock setting menu, firstly the renewed time and date would be displayed on the
screen, then it returns to the main menu; if you didn't reset the time and date , when exiting the clock
setting menu, the device would return to the main menu directly.
g) Exit the main menu
On the main menu interface, press the "menu button" to exit the main menu.
F. PC SOFTWARE OPERATION
Please connect the device with computer by the USB data line, then double click "SpO
icon to run the PC software. The functions such as uploading data and change device ID could be
carried out by the software. Please refer to <SpO
If the users choose to turn on the synchronizing display function on computer, it would probably
take several seconds for the data to appear on the computer screen. (If there is no data on the
computer screen, unplug USB data line, then repeat step "E" again).
6.2 Attention for operation
A. Please check the device before using, and confirm that it can work normally.
B. The finger should be in a proper position (see the attached illustration of figure 1 for reference), or
else it may result in inaccurate measure.
C. The SpO
sensor and photoelectric receiving tube should be arranged in a way with the subject's
2
arteriole in a position there between.
D. The SpO
sensor should not be used at a location or limb tied with arterial canal or blood pressure
2
cuff or receiving intravenous injection.
E. Do not fix the SpO
measure of SpO
and pulse rate.
2
F. Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light,
infrared heater, direct sunlight and etc.
G. Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
H. Testee can not use enamel or other makeup.
I. Please clean and disinfect the device after operating according to the User Manual (7.1).
6.3 Clinical restrictions
A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject is
required. For a subject with weak pulse due to shock, low ambient/body temperature, major bleed-
ing, or use of vascular contracting drug, the SpO
measurement will be more sensitive to interference.
B. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green
and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosali-
cylic hemoglobin, and some with icterus problem, the SpO
inaccurate.
C. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor
blamed for serious error of SpO
D. As the SpO
value serves as a reference value for judgement of anemic anoxia and toxic anoxia,
2
some patients with serious anemia may also report good SpO
7. MAINTAIN, TRANSPORTATION AND STORAGE
7.1 Cleaning and disinfecting
Using medical alcohol to disinfect the device, nature dry or clean it with clean soft cloth.
7.2 Maintain
A. Please clean and disinfect the device before using according to the User Manual (7.1).
B. Please change the battery when the screen shows
C. Take out the battery if leave the equipment unused for long time.
D. The device needs to be calibrated once a year (or according to the calibrating program of hospital).
It also can be performed at the state-appointed agent or just contact us for calibration.
sensor with adhesive or else it may result in venous pulsation and inaccurate
2
measure.
2
ENGLISH
Assistant user manual> for details.
2
waveform (PLETH) will decrease. In this case, the
2
determination by this monitor may be
2
measurement.
2
.
Assistant"
2