Instruction description:
u Key operation:
1. Short pressing "
2. Short pressing "
3. Short pressing "
u Touchscreen operation:
1. Pressing the corresponding parameter button on the lower screen to directly enter into corresponding
functional parameter setup screen.
2. Pressing "
" button on the lower screen to exit from Setup Menu Screen.
Note: the device will save the latest setup settings automatically and the most of saved settings are non-volatile,
that's to say, when you shut down the device and power up it next time, every setting item shows the settings
saved last time except the items like ECG cable setting and NIBP working mode.
4.9.1 ECG Setup (Optional)
Screen description:
"HR alarm high": high alarm limit setting for HR. Setting range: 1~350, the factory default is 180.
²
"HR alarm low": low alarm limit setting for HR. Setting range: 0~349, the factory default is 40.
²
"Unfilter": the switch of filter mode. Select "On" means enhance mode for the filter with extended
²
bandwidth (0.05Hz~40Hz), select "Off" means filter with normal bandwidth (0.5Hz ~ 40Hz).
"Gain": set ECG waveform gain, "X1, X2, X4, X1/2, X1/4" for optional.
²
"X1": Waveform scale with base gain
"X2": Twice scale size of the base gain.
"X4": 4 times scale size of the base gain.
"X1/4": 1/4 size of the base gain
"X1/2": Half scale size of the base gain
User Manual for Vital Signs Monitor
" key or "
" key to shift cursor to corresponding functional group setting.
" key to confirm and enter into corresponding functional parameter setup screen.
" key to exit from Setup Menu Screen.
Figure 4.11A ECG setup
35