Clinical Safety; Electromagnetic Compatibility (Emc) - FujiFilm Sonosite S II Stand Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Cautions:

Clinical safety

WARNING:

Electromagnetic compatibility (EMC)

See the safety section of the ultrasound system user guide and supplemental guides for information on
electromagnetic compatibility with IEC 60601-1-2, including the Manufacturer's Declaration.
14
Excessive bending or twisting of cables can cause a failure or intermittent
operation.
Improper cleaning or disinfecting of any part of the accessories and peripherals
can cause permanent damage. For cleaning and disinfecting instructions, see
"Cleaning and disinfecting"
instructions.
Use only peripherals recommended by FUJIFILM SonoSite.
Accessible metal on the signal inputs and outputs on the back and side of the
ultrasound system is not protectively earthed. Do not perform a high current
ground impedance test involving this part.
To avoid misdiagnosis, do not use non-medical (commercial) grade peripheral
monitors. These monitors have not been verified or validated by FUJIFILM
SonoSite as being suitable for diagnosis.
on page 11 and the manufacturer's operating

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido