Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic
field
IEC61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
<5% U
(>95% dip
T
in U
) for 0.5 cycle
T
40% U
(60% dip
T
in U
) for 5 cycles
T
70% U
(30% dip
T
in U
) for 25 cycles
T
<5% U
(>95% dip
T
in U
) for 5 sec
T
3A/m
ENGLISH
<5% U
(>95% dip
T
in U
) for 0.5 cycle
T
40% U
(60% dip
T
in U
) for 5 cycles
T
70% U
(30% dip
T
in U
) for 25 cycles
T
<5% U
(>95% dip
T
in U
) for 5 sec
T
3A/m
Mains power quality
should be that of a typical
commercial or hospital
environment. The device can
continue the operation during
power mains interruptions
due to the usage of battery.
Mains power quality
should be that of a typical
commercial or hospital
environment.