Operation
User navigation
Refer to the controller documentation for keypad description and
navigation information.
Configure the sensor
Use the Configure menu to enter identification information for the sensor
and to change options for data handling and storage.
1. Push the MENU key and select Sensor Setup, [Select Sensor],
Configure.
2. Use the arrow keys to select an option and push ENTER. To enter
numbers, characters or punctuation, push and hold the UP or DOWN
arrow keys. Push the RIGHT arrow key to advance to the next space.
Option
Description
EDIT NAME
Changes the name that corresponds to the sensor on
the top of the measure screen. The name is limited to
10 characters in any combination of letters, numbers,
spaces or punctuation.
SENSOR S/N
Allows the user to enter the serial number of the sensor,
limited to 16 characters in any combination of letters,
numbers, spaces or punctuation.
SELECT MEASURE Changes the measured parameter to conductivity
(default), TDS (total dissolved solids), salinity or
resistivity. When the parameter is changed, all other
configured settings are reset to the default values.
DISPLAY FORMAT
Changes the number of decimal places that are shown
on the measure screen to auto (default), X.XXX, XX.XX,
XXX.X or XXXX. When set to auto, the number of
decimal places changes automatically with changes in
the measured value.
MEAS UNITS
Changes the units for the selected measurement—
conductivity: µS/cm (default), mS/cm, µS/m, mS/m or
S/m; TDS: ppm (default) or ppb; salinity: ppt (default);
resistivity: MΩ (default) or kΩ
10 English
Option
Description
TEMP UNITS
Sets the temperature units to °C (default) or °F
T-COMPENSATION Adds a temperature-dependent correction to the
measured value—linear (default: 2.0%/°C, 25 °C),
ammonia, temp table (enter x,y points in ascending
order), none, natural water or pure water. For special
applications, a user-defined linear compensation can
be entered (0–4%/°C, 0–200 °C). Natural water
compensation is not available for TDS.
CONFIG TDS
TDS only—changes the factor that is used to convert
conductivity to TDS: NaCl (default, 0.49 ppm/µS) or
custom (enter factor between 0.01 and 99.99 ppm/µS).
CELL CONSTANT
Sets the cell constant range to 0.05, 0.5, 1.0 (default),
5.0 or 10.0. After the range is selected, the user can
enter the certified K value from the label on the sensor
cable. When the certified K value is entered, the
calibration curve is defined.
CABLE LENGTH
Sets the actual length of the sensor cable to improve
measurement accuracy (default: 20 ft (Polymetron
sensors default: 5 ft)).
TEMP ELEMENT
Sets the temperature element for automatic
temperature compensation to PT100 or PT1000
(default). After selection, the user should enter the
certified T-factor from the label on the sensor cable for
best accuracy. If no element is used, the type can be
set to manual and a value for temperature
compensation can be entered (manual default: 25 °C).
Note: If a sensor with a PT100 or PT1000 element is
set to manual and the sensor is replaced or the sensor
days are reset, the TEMP ELEMENT automatically
changes to the default setting.
FILTER
Sets a time constant to increase signal stability. The
time constant calculates the average value during a
specified time—0 (no effect, default) to 60 seconds
(average of signal value for 60 seconds). The filter
increases the time for the sensor signal to respond to
actual changes in the process.