Sensor diagnostic and test menu
The sensor diagnostic and test menu shows current and historical
information about the instrument. Refer to
diagnostic and test menu, push the MENU key and select Sensor Setup,
[Select Sensor], DIAG/TEST.
Table 5 Sensor DIAG/TEST menu
Option
Description
SENSOR INFORMATION Shows the name and serial number that was entered
by the user.
CARD INFORMATION
Shows the version and the serial number for the
sensor module.
CAL DAYS
Shows the number of days since the last calibration.
CAL HISTORY
Shows a list of and details for each calibration.
RESET CAL HISTORY
Resets the calibration history for the sensor (requires
service-level passcode). All previous calibration data
is lost.
SENSOR SIGNALS
Shows the current sensor signal and span in µS/cm.
SENSOR DAYS
Shows the number of days that the sensor has been
in operation.
RESET SENSOR DAYS
Resets the number of days that the sensor has been
in operation.
Error list
Errors may occur for various reasons. The reading on the measurement
screen flashes. All outputs are held when specified in the controller menu.
To show the sensor errors, press the MENU key and select Sensor Diag,
[Select Sensor], Error List. A list of possible errors is shown in
16 English
Table
5. To access the sensor
Table
6.
Table 6 Error list for conductivity sensors
Error
Description
MEAS TOO
The measured value is >
HIGH
2,000,000 µS/cm,
1,000,000 ppm or 20,000
ppt
MEAS TOO
The measured value is < 0
LOW
µS/cm, 0 ppm or 0 ppt
ZERO TOO
The zero calibration value
HIGH
is > 500,000 counts
ZERO TOO
The zero calibration value
LOW
is < –500,000 counts
TEMP TOO
The measured
HIGH
temperature is > 130 °C
TEMP TOO
The measured
LOW
temperature is < –10 °C
ADC
The analog to digital
FAILURE
conversion failed
SENSOR
The sensor is missing or
MISSING
disconnected
SENS OUT
The sensor signal is
RANGE
outside of the accepted
limits for the cell constant
that is used (0.01 and 0.05:
100 µS/cm; 0.5: 1000 µS/
cm; 1: 2000 µS/cm; 5:
10,000 µS/cm; 10:
200,000 µS/cm)
Resolution
Make sure that the sensor is
configured for the correct cell
constant.
Make sure that the sensor is
configured for the correct cell
constant.
Make sure that the sensor is held in
air during zero calibration and is not
located near radio frequency or
electromagnetic interference. Make
sure that the cable is shielded by
metal conduit.
Make sure that the sensor is
configured for the correct
temperature element. Refer to
Test
the conductivity sensor
on page 15.
Make sure that the sensor module is
fully inserted into the controller
connector. Replace the sensor
module.
Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.
Make sure that the sensor is
configured for the correct cell
constant.