Section B: Surgical Procedure; Precautions; Irrigation - Codman CERTAS Plus Manual Del Usario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 143
9. Return all tools to the storage case to prevent damage.
Note: fully collapse the Adjustable Height Locator
Tool (See Figure 2B).

SECTION B: Surgical Procedure

Precautions

Silicone has a low cut and tear resistance; therefore,
exercise care when placing ligatures so as not to tie
them too tightly. The use of stainless steel ligatures
on silicone rubber is not recommended.
Do not use sharp instruments when handling the
silicone valve or catheter; use shod forceps. Cuts or
abrasions from sharp instruments might rupture or
tear the silicone components.
Do not fold or bend the valve during insertion.
Folding or bending might cause rupture of the
silicone housing, needle guard dislodgement,
or occlusion of the fluid pathway.
Verify proper placement and integrity of ligatures
at all tubing junctions to prevent obstruction of
the catheter lumen and tears or abrasions of the
silicone tubing.
Exercise extreme care to prevent the silicone
components of the system from coming in contact
with towels, drapes, talc, or any linty or granular
surfaces. Silicone rubber is highly electrostatic and,
as a result, attracts airborne particles and surface
contaminants that could produce tissue reaction.
Do not immerse the BACTISEAL Catheter in
antibiotic solutions. Only use sterile water or normal
saline to immerse the BACTISEAL catheter as other
solutions are not recommended and may cause
precipitation and consequent adverse effects (e.g.,
catheter occlusion, hydrocephalus). Keep the time
the BACTISEAL catheter is immersed in sterile
water or normal saline to a minimum (i.e., seconds)
to reduce the risk of introducing infectious agents.
A pale orange color might be imparted to the
immersion solution. Manometer testing of the valve
with BACTISEAL Catheter is not advised.

Irrigation

It is required that the valve be irrigated before
implantation to help ensure proper performance.
1. Hold the valve vertically with the distal connector
barb pointing downward.
2. Using a syringe, slowly and gently fill the entire valve
system with pyrogen-free, sterile saline solution or,
for catheters other than BACTISEAL Catheters, an
appropriate antibiotic solution. Once fluid begins
flowing through the valve mechanism, the distal
connector barb of the valve can be pointed upward to
assist in evacuating air from the system.
Note: A priming adapter with inlet tubing is provided
to facilitate irrigation.
CAUTION: Do not fill, flush, or pump the valve with
fluid in which cotton, gauze, or other lint-releasing
material has been soaked.
Note: More pressure might be required on the syringe
before fluid starts flowing through the valve mechanism.
This is normal and will only occur during initial irrigation
of the valve. A popping sound may occur.
13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Certas

Tabla de contenido