On the milling side, make sure the shank of another scanner
moves smoothly, and insert it inversely into the hand piece of
the milling side.
Open the screw at the spherical shell table CL .
Carefully insert the inverted scanner into the calibration hole
on the scanner side. On the milling side, the scanner is above
the hole since initially the scanner was moved out by 2 mm.
38
With the scanner inserted on the lefthand side, carefully
move the table so that on the righthand side the scanner
slides into the hole.
Carefully move the table so that the right scanner also slides
into the hole. It is sufficient if about 1/3 of the scanner sink
into the hole.
CaUtIOn! Du not move the table when both scanners are in
their holes. This might damage the micromotor!!!
Lift milling guide, take out scanner.