1. Check whether the instrument is clean after conditioning under good
lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If it is still dirty, repeat the process.
 For further conditioning, the instrument should be dry and free of
residue.
3. Blow the instrument out with max. 3 bar.
4. Manual care of mechanical parts [ → 33].
5. Manual care of the push button chuck [ → 35].
6. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Sterilize the instrument [ → 36].
7.3

Manual cleaning and disinfection

NOTICE! Condition immediately, or at the latest, one hour after
Immediately after treatment
treatment.
7 Conditioning
31