B
Option
Clean and disinfect system and transducer to a low level (non-critical
uses)
Use the following procedure to clean and disinfect the ultrasound system and transducer if it has not come
into contact with blood, broken skin, mucosal membranes, or bodily fluids.
WARNING
Follow the manufacturer's instructions when using cleaners and disinfectants. The cleaners and disinfectants
listed in the procedure are both chemically compatible and have been tested for efficacy with the system
and transducers. Confirm that the cleaners and disinfectants are appropriate for your facility's use.
WARNINGS
Cautions
To clean and disinfect the system and transducer
1 Turn off the system by pressing the Power button.
2 Unplug the power cord from the outlet.
Cleaning and disinfecting
If the system or transducer has come into contact with any of the following,
use the high-level cleaning and disinfection procedure. See
disinfect system and transducer to a high level (semi-critical uses)"
page 10:
Blood
Broken skin
Mucosal membranes
Bodily fluids
To avoid electrical shock, before cleaning, disconnect the system from the
power supply.
Wear the appropriate personal protective equipment (PPE) recommended by
the chemical manufacturer, such as eyewear and gloves.
Do not spray cleaners or disinfectants directly on the system surfaces or on
system and transducer connectors. Doing so may cause solution to leak into
the system, damaging it and voiding the warranty.
Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using
a non- recommended disinfecting solution or incorrect solution strength can
damage the System and transducer and void the warranty. Follow the
disinfectant manufacturer's recommendations for solutions strengths.
Do not attempt to disinfect a transducer or transducer cable using a method or
chemical that is not included here. Doing so can damager the transducer and
void the warranty.
"Clean and
on
15