Complete / Stopped – 75 minutes has elapsed (safety feature)
Timer shows flashing 75
Status Indicator light is steady green
Vacuum and heater off
LEDs on in vacuum head
Vacuum Fault: – vacuum seal has been compromised (or never made)
Timer flashes running time elapsed (not counting)
Status Indicator light flashes amber after ten seconds and the
Vacuum and heater off
LEDs off in vacuum head
Other Fault / Service Unit: – heater broken, tubing unplugged, any problem other than vacuum
Timer shows two underscores (_ _)
Status Indicator light flashes amber and the
Vacuum and heater off
LEDs off in vacuum head
CONTROL UNIT STAND ASSEMBLY INSTRUCTIONS
Base Pole

TOOLS REQUIRED:

1/2" Box Wrench
1. Insert base pole with tapered bushing into top of base.
2. Screw hex head bolt with washers into base pole as shown.
3. Tighten securely with 1/2" box wrench.
Base
service icon
Tapered Bushing
Fender Washer
Lock Washer
Hex Head Bolt