Rated maximum output power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated by
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts [W]
according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
English
Fluent Fluid Management System
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2√P
d = 1.2√P
0.12
0.38
1.2
3.8
12
Chapter 13: Electromagnetic Compatibility
800 MHz to 2.7 GHz
d = 2.3√P
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23