This chapter lists the technical specifications for the Fluent Fluid Management System and information about power cord safety.
Technical Specifications
Table 1 Fluent Fluid Management System Technical Specifications
Item
Model or type designation
Mains voltage range [V]
Supply Frequency Range [Hz]
Fuse designation
Mode of Operation
Power Consumption
Upper Voltage Range
Normal Operation
Peak
Lower Voltage Range
Normal Operation
Peak
Protection class (I, II, III)
Application part type (B, BF, CF)
Defibrillator protected (yes, no)
Protection type (IP code)
Classification (I, IIa, IIb, III acc. To Appendix IX of European
MDD)
Conformity with the following standards
Operating conditions
Use possible with flammable anesthetic gases
Storage and transportation conditions
Maximum sound level
Adjustable Values
Pressure range (mmHg)
Deficit Limit (mL)
Measurement range
Flow (mL/min)
Deficit (mL)
Accuracy
Pressure (mmHg)
Flow (mL/min)
Deficit (mL)
English
Fluent Fluid Management System
Chapter 12: Technical Specifications
Specification
FLT-100
100-240 VAC
50-60Hz
T5AH, 250 V fuses
Continuous
350W
240 VAC
264 VAC
120 VAC
90 VAC
I
BF
No
IP21
IIb
IEC 60601-1:2005 + A1:2012
IEC 60601-1-6
IEC 62366-1:2015Ed.1
IEC 62304:2006 Ed.1 +A1
IEC 60601-1-2:2014
15-30ºC / 58-85ºF
20-80% rel. humidity, Non-Condensing
8000 ft (2438.4 Meters) max. altitude above sea level for use
No
-10-60ºC / 14-140ºF
10-80% rel. humidity, Non Condensing
≤75 decibels (dbA at 1meter)
40 - 120
100 - 2500
0-650
-9999 / +99999
± 15
± 50
± 50 under normal use
Chapter 12: Technical Specifications
43