Maintenance (Fuse Replacement); Check Before Each Use - DJO ARTROMOT-K1 classic Instrucciones De Uso

Ocultar thumbs Ver también para ARTROMOT-K1 classic:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 76
6.2 Maintenance
(fuse replacement)

Check before each use

Visually inspect the device for signs of
mechanical damage before each use.
If you detect damage or malfunctions that
may impair the safety of the patient or of
the operator, have the device repaired
before using it.
Technical inspections
Due to their design, the bearings and joints
of the ARTROMOT®-K1 classic do not
require maintenance and all materials are
protected from corrosion, BUT: For safety,
the devices require regular maintenance.
To maintain the functional and operational
safety, check all components for damage
and loose connections at least once a year.
These checks should be performed by
persons with adequate training and expe-
rience. Damaged and worn parts must im-
mediately be replaced with original spare
parts by authorized staff.
Warning!
m
Patient hazard –
equipment malfunction and damage
− Refer repair and maintenance to
authorized persons. Persons are
authorized after training by a special-
ist trained and commissioned by the
manufacturer.
ORMED GmbH will make all docu-
ments required for servicing, such as
circuit diagrams, parts lists, descrip-
tions or calibration instructions, avail-
able to authorized experts.
The inspections can be carried out by DJO
Technical Service within the framework
of a service agreement. Contact DJO for
details.
Other than that, the manufacturer
does not require any other regular
maintenance.
Note!
Regarding further technical or other in-
spections or inspection intervals, observe
applicable local requirements, such as
IEC 62353, DGUV3 or similar regulations
to be observed by users of medical
equipment or electric devices.
Fuse replacement
Warning!
m
Patient hazard –
equipment malfunction and damage
The replacement of fuses must be
referred to specialists as defined in
DIN VDE 0105 or IEC 60364 or other
applicable standards (e.g. biomedical
technicians, electricians, electronics
installers).
Fuses used must be
S# < 20,000: T1A L250Vac
S# > 20,000: T2A H250Vac.
40

Publicidad