Care And Cleaning - arjo TheraRest Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 87
EN

Care and Cleaning

The following process is recommended, but should be adapted to comply
with local institution protocols. If you are uncertain, you should seek advice
from your local Infection Control Specialist. TheraRest MRS should be
routinely decontaminated between patients and at regular intervals while in
use.
CAUTION
Do not use Phenol-based solutions or abrasive componds or
pads on the cover during decontamination process as these will
damage the surface coating. Do not boil or autoclave the cover.
Covers should be wiped down immediately after it has been
Fire Barrier
removed. It is recommended that Personal Protective Equipment (PPE) is
contact with loose particles. To ensure optimal product performance, avoid
be replaced if soiled or damaged.
Cover Cleaning Options
Sewn TheraRest MRS (Top cover cannot be removed from base)
1. Remove or push the bed linens to the center of the mattress.
CAUTION
The sewn TheraRest MRS cover is cleaned by wipe down
methods only. Do not launder as damage to the base may occur.
2. Wipe and rinse any soiling from the mattress surface and base. Use
1000 ppm chlorine or 70% alcohol.
3. Rinse with a wet towel to remove chemicals.
12

Publicidad

Tabla de contenido
loading

Tabla de contenido