These oxygen cylinder holders are suitable for use with all Arjo Enterprise® and
Citadel
The table shows the range of cylinder sizes that can be used with each model
of holder. The maximum permitted cylinder height is 870mm (including valve).
Refer to the bed instructions for use to confirm compatibility before using this accessory.
If a serious incident occurs in relation to this medical device, affecting the user, or the
patient then the user or patient should report the serious incident to the medical device
manufacturer or the distributor. In the European Union, the user should also report the
serious incident to the Competent Authority in the member state where they are located.
Installation and use
1. Fit the oxygen cylinder holder into either of the head end sockets (1).
2. Rotate the accessory so that the notch (2) locates over the pin at the bottom of the socket.
3. Place a suitably sized oxygen cylinder in the holder and tighten the clamp screw (3) to
secure it.
4. The oxygen cylinder holder has an accessory socket (4) for an IV pole.
The oxygen cylinder holder can be positioned in two ways:
Outwardly, perpendicular from the head end of the bed.
Outwardly, perpendicular from the side of the bed.
TM
Beds.
Model no.
ENT-ACC08
ENT-ACC18
ENT-ACC58
2
Cylinder diameter d
Minimum
100 mm
140 mm
119 mm
WARNING
3
1
EN
Maximum
108 mm
151 mm
122 mm
4