Attaching the Nosecone and the Flue
1. Slide the larger end of the nosecone over the tip
and slide it down over the neck of the handpiece
making sure the white dots are aligned.
5
2. Rotate the nosecone clockwise until it locks into
place and the white dot on the nosecone aligns with
the white dot on the body of the handpiece.
Make sure the white dots are aligned on the nosecone
and handpiece.
5
3. Slide the flue onto the nosecone base.
4. Make sure that the flue is aligned with the pre-
aspiration holes near the tip.
NOTE:
If the aspiration holes are completely covered, most
of the irrigation fluid may be aspirated back through
the handpiece, impairing cavitation, thus reducing
Surgical Aspirator efficacy.
Attaching the CUSA Quick Connect
Tubing
1. Open the CUSA Quick Connect
tubing set using proper sterile technique.
2. Pick up the larger loop of the irrigation and
aspiration tubing set to be brought into the sterile
field and assembled to the handpiece.
3. Place the cartridge and smaller loop of the irrigation
and aspiration tubing set aside until it is time to
connect it to the console.
4. Attach the aspiration tubing to the aspiration port at
the base of the handpiece.
4
cartridge and