The display should read SET LOADED and will show one of the following opening menus, depending on the type of pump set
detected, and whether continuous or intermittent feeding mode has been selected.
Figure 7A. Opening menu for
Continuous Feed mode with a
feed-only set.

Prime Pump

The Kangaroo™ ePump™ enteral feed and flush pump may be primed automatically with a single button push, including the flushing
line if a feed & flush set is loaded. The pump can also be primed in a more interactive method using the hold-to-prime feature. The
flow rate for priming, whether Auto Prime or Hold-to-Prime, is 1960 ml/hr (32.7 mL/minute). Note: During priming the Flow Error
alarm is disabled.
After a pump set has been loaded and the status line shows SET LOADED, press
to get one of the PRIME PUMP menus shown in Figures 8A-8B.

Auto Priming

The AUTO PRIME option will be available if the pump senses that the currently installed pump set was not previously primed, and
that there is no fluid in the line near the rotor. If auto-priming is desired, press and release
the line(s). For feed & flush pump sets, the auto prime feature will automatically prime both lines, starting with the flush line.
Press
STOP to cancel AUTO PRIME.
4
When auto-priming is completed, the status line will show AUTO PRIME COMPLETE and will no longer show the AUTO PRIME
option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete the priming, as described below.

Hold-to-Prime

The hold-to-prime options allow for precise interactive control of the priming process.

Feed-Only Pump Sets

For feed-only pump sets, the screen will appear as in Figure 8A. Press and hold
primed down to the stepped end connector at the end of the pump set.
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Figure 7B. Opening menu for
Continuous Feed mode with a feed
& flush set.
Figure 8A. PRIME PUMP menu for
feed-only set.
Figure 7C. Opening menu for
Intermittent Feed mode with a
feed-only set.
PRIME PUMP in the opening menu (Figures 7A-7D)
4
Figure 8B. PRIME PUMP menu for
feed & flush set.
4
HOLD TO PRIME FEED until the feed line has been
4
11
Figure 7D. Opening menu for
Intermittent Feed mode with a
feed & flush set.
AUTO PRIME to automatically prime
Table of Contents

Publicidad

Tabla de contenido
loading