Flush Now; More Options; Buzzer Alarm Volume; History - Coviden Kangaroo ePump Manual De Instrucciones

Bomba para irrigación y alimentación enteral con abrazadera para palo de gotero, programable
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 197

Flush Now

(Not Available in EZMODE)
During the RUNNING mode, the pump can be diverted to immediately perform a flush (flush on demand) by pressing
NOW in the RUNNING menu, input a desired flush volume, then press
the screen will prompt for the volume to be immediately flushed, and will show the previously programmed flush volume as default
but may be adjusted to the users desired volume. The screen will appear as the standard flush input screen as shown in Figure 12B
(or 14B). Setting the FLUSH NOW flushing volume is temporary and will not change the main/periodic flushing volume previously
programmed. If
FLUSH NOW is pressed by accident; the user may either press the cancel button or allow the screen to return to
4
the RUNNING screen after 10 seconds, if there is no input for the flush volume.
If stopping a FLUSH NOW before the set volume has been delivered is desired, press
press
ENTER. The FLUSH NOW will stop immediately and the pump will return to the RUNNING mode.
4
Figure 19. MORE OPTIONS menu.

History

(Not Available in EZMODE)
Press
HISTORY in the MORE OPTIONS menu to access the HISTORY screen. Press
4
buttons in Figure 20 to set the time history of interest. The totals of FEED mL and
the
4
FLUSH mL will be shown for the requested history time.
Warning: A maximum of 72 hours of past history can be retrieved, excluding data from the
current hour. Note: The 72-hour history includes time when the pump is powered OFF.

Language Selection

Press
LANGUAGES in the MORE OPTIONS menu to change the preferred language for the
4
screen interface.

Continuous/Intermittent Mode

Press
CONTINUOUS/INTERMITTENT in the MORE OPTIONS menu to access the SELECT MODE menu, Figure 10B. Use this menu
4
to select between continuous feeding mode and intermittent feeding mode. See subsection "Selecting Continuous or Intermittent
Feed Mode" for more information.
DONE in the MORE OPTIONS menu to exit the menu.
Press
4
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp

More Options

MORE in one of the opening menus, Figures 5A-5B or Figures 7A-7D, to access the MORE
Press
4
OPTIONS screen, Figure 19.

Buzzer Alarm Volume

Press
BUZZER in the MORE OPTIONS menu to access the BUZZER screen and increase or
4
decrease the alarm buzzer volume. The alarm will be audible as the volume setting is changed.
This sets the volume level for any warning or error that may occur with the pump. High Priority
alarms, such as System Error, Rotor Error, and Pump Set Dislodged will always use the maximum
volume setting and will be louder than other alarms.
(Not Available in EZMODE)
19
ENTER. See Figure 15A. When
4
FLUSH NOW and enter a volume of zero and
4
FLUSH
4
FLUSH NOW is pressed,
4
Figure 20. Feed and Flush totals for
up to 72 hours of previous history,
excluding data from the current
hour.
Table of Contents

Publicidad

Tabla de contenido
loading

Tabla de contenido