Section II — Safety and Warnings
Caution: U.S. federal law restricts the sale of this device to physicians or to their direct representatives
1. Read this booklet thoroughly before using the Kangaroo™ ePump™ enteral feed and flush pump.
2. Do not use this device near flammable anesthetics.
3. Disconnect power supply before cleaning or servicing.
4. Use only Kangaroo™ ePump™ Enteral Feed and Flush Pump A/C Power Adapter with pump. See Section XIII - Service Part
Numbers for replacement of power adapter. Use of an alternate consumer style A/C power adapter may cause damage
to the charging circuit and battery of the Kangaroo™ ePump™ feeding pump.
5. Danger: Use only Kangaroo™ ePump™ enteral feed or feed with flush pump sets with this device. Pump is not compatible with
other pump sets. Use of other feeding sets with this pump can create hazardous situations, including free-flow conditions that
can result in overfeeding, underfeeding, formula in the lungs, and death to a patient.
6. For performance tests, see Section VI - Performance Tests. For other integrity checks, consult with a qualified biomedical
technician or contact the manufacturer (Section XI - Customer Service).
7. For service or for technical information, please contact Customer Service (Section XI).
8. Do not open the main housing, as there are no user-serviceable parts inside. Opening may affect function of device and voids
the warranty. To replace battery, see Section VIII.
9. Dispose of old battery-powered electronic equipment in a manner consistent with institutional policy for expired
equipment disposal.
10. Cleaning frequency and practices must be consistent with institutional policy for cleaning of non-sterile devices. See Section VII
- Cleaning, for instructions on cleaning the Kangaroo™ ePump™ enteral feed and flush pump.
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Feeding Set Capture