Caregiver Alarm Notification
All alarms are intended to be heard by operators that are within hearing range of the pump buzzer. The pump buzzer is located on
the back of the pump. The pump is designed so the alarm can be heard within the patient's room, at a minimum. The display and
LED alarm indicators are intended to be seen by an operator within the room, facing the front of the pump. Since audible alarms are
limited by distance, it is recommended that the operator conduct a check to determine at what distance the alarm can still be heard.
Note:
Going outside of the patient's room may make hearing the alarms more difficult.
The Error, Warning, and Information Screens are described below:

System Error

(High Priority Alarm)
The system error screen is the most general form of error, Figure 25. The only way to exit from
a System Error is to power down. No CONTINUE option is allowed because of the severity
of the error. An error number is also displayed on the screen, for reference purposes. This
number should be reported when calling the Customer Service Support Line. Detection of this
alarm condition may take up to 1 minute to occur under normal operating conditions.
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
26
Figure 25. General System Error
screen. See Customer Service
section.