Verifying functionality of the alarm system
The System Performance Test is sufficient to verify proper pump function. However, another quick test that can be performed to
confirm audibility and function of alarms is:
1. Load a new feeding set onto the pump
2. Leave the feeding set empty!
3. Run Auto Prime
4. Once the pump begins priming, it will issue a Feed or Flush Error alarm
5. Confirm that the audible alarm and colored LEDs all properly indicate a Feed or Flush Error
Caution: To allow proper operation of audible alarms, avoid blocking or obstructing the series of buzzer holes located
in the back of the pump.
Section VII — Cleaning
Cleaning should be performed as needed. It may also be desirable to define cleaning intervals based on knowledge of the
environment in which the pump is used. Only personnel trained in the cleaning of medical devices should perform cleaning.
Caution: Do not immerse pump or A/C power adapter in water or other cleaning solution; clean using a damp (not
wet) cloth or sponge. Failure to follow the cleaning procedures described herein could result in hazards to users.
As with any AC powered electrical device, care must be taken to prevent liquid from entering the pump to avoid
electrical shock hazard, fire hazard, or damage to electrical components.
If any of the following events occur, do not use the pump until it has been properly cleaned and serviced by personnel trained in
servicing Kangaroo™ ePump™ enteral feed and flush pumps:
• wetting of the pump's A/C power adapter or leakage into the pump interior during cleaning.
• spillage of large amounts of formula onto the pump exterior, or any spillage onto the A/C power adapter.
General Cleaning Directions
Cleaning of Kangaroo™ ePump™ enteral feed and flush pumps must be performed as follows:
Caution: Disconnect pump from AC power source before cleaning. After cleaning, do not connect to AC power source
until pump and A/C power adapter are thoroughly dry.
• A mild detergent should be used for general cleaning. For disinfection a 10% hypochlorite solution, isopropyl alcohol or
ammonium chloride based solutions may be used. Allow excess moisture to evaporate from the cord prior to use of A/C
Power Adapter.
• Do not use strong cleaners such as Spray Nine™*, pHisoHex™*, Hibiclens™*, or Vesta-Syde™* or cleaners that may contain
Ethoxylated C9-C11 alcohols, 2-Butoxyethanol, Hexachlorophene, Chlorhedidine Gluconate, Subtilisins, Sodium Tetraborate
Decahydrate, Triethanolamine, or other similar ingredients because damage to the pump case housing can result .
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
22