2. Description of device
2.1 Intended use
EN
10
Description of device
MEDUMAT Easy CPR is an automatic oxygen respira-
tion device (short-term ventilator) with additional in-
halation facility.
You can use MEDUMAT Easy CPR:
•
to revive patients at the site of the emergency;
•
for longer periods in more protracted emergen-
cies, e.g. fires;
•
for short-term O
mask.
You can use MEDUMAT Easy CPR while transporting
patients:
•
between the various rooms and departments of
a hospital;
•
between the hospital and other premises;
•
in emergencies;
•
when transport over considerable distances is
planned.
MEDUMAT Easy CPR:
•
is designed to provide controlled ventilation to
persons of 10 kg body weight or more;
•
is used to treat respiratory arrest;
•
can be preset to parameters that ensure evenly
balanced ventilation, provided that the selected
maximum ventilation pressure P
exceeded.
inhalation using a respiration
2
max
is not