5.8 CPR mode
60
50
40
CP R
30
20
10
0
Automatic ventilation with the selected frequency is
stopped when CPR mode is activated. With the help
of MEDUtrigger, you can trigger single respiratory
strokes with the selected tidal volume. This enables
you to choose the respiratory frequency adminis-
tered yourself.
Note:
CPR mode can only be activated if the MEDUtrigger
is connected.
1. Switch on CPR mode by pressing the CPR key.
The LEDs on the CPR key and MEDUtrigger light
up to indicate operating readiness. If audio re-
CPR
sponse is enabled, you should hear the message
"CPR mode activated! Ventilation triggered
manually!".
2. If metronome is enabled, you should hear the
message "Perform chest compressions now!".
Undertake chest compressions to the beat of the
metronome 30 times (110 min
rises for the last three strokes of the metronome.
3. You should hear the message "Provide 2
ventilations now!". You now have 5 seconds to
trigger 2 ventilation strokes.
Ventilate the patient by pressing the
MEDUtrigger key:
–
or
–
A ventilation stroke cannot be triggered during
inspiration and expiration phases. The LEDs on
the MEDUtrigger do not light up at this time.
Hold this key until the second ventilation
stroke has begun,
manually trigger the second ventilation
stroke by pressing the MEDUtrigger key
again, if the expiration phase of the first
ventilation stroke has ended.
Using the ventilator
-1
). The tone pitch
EN
29