Functional check
Checking the tidal volume
1. MEDUMAT Easy CPR must be switched off and
the oxygen cylinder must be open.
2. Attach the test bag to the patient valve with the
adapter from test set WM 15357.
3. Select the following settings:
Note
During the expiration phase, you must simulate
the expiration stroke of the test bag by hand. To
do so, place the test bag on a firm base. During
the expiration phase, press the test bag with the
flat of your hand until the volume is completely ex-
pelled through the patient valve.
4. Switch on MEDUMAT Easy CPR. The test bag
must become fully inflated during inspiration.
This ensures a tidal volume of 950 ml per
inspiration stroke. At all events the test bag is
not sufficiently inflated if a disconnection alarm
is triggered.
Note
In combination with test bag WM 1454, these
settings may lead to a
bag inflates fully, the functional check has been
passed.
5. Switch off MEDUMAT Easy CPR again.
Risk of injury if test bag removed
incorrectly!
If the test bag is removed incorrectly, the con-
nector of the test bag may remain on the pa-
tient's hose system. The increased airway resis-
tance on inspiration which results may injure the
patient.
V
: 950 ml/ Frequency: 10 min
t
P
:
(45 mbar)
max
When removing the test bag, always take it
off at the connector.
-1
Stenosis
alarm. If the test