Important
In this test the rise in pres-
sure is so strong that the
pressure gauge needle may
overswing into the red
zone. There are technical
reasons for this, and it does
not indicate any malfunc-
tion.
EN
56
Functional check
Disconnection (interruption of
breathing system)
1. Open the oxygen cylinder.
2. If necessary: Remove the tube or the ventilation
mask from the patient valve.
3. Switch on MEDUMAT Easy CPR.
4. Switch the mask/tube switch to
5. Keep the ventilation connector on the patient
valve closed with the flat of your hand during
two successive inspiration phases. The
alarm should be triggered.
If audio response is enabled, the ventilator
announces "Check airways and settings!".
6. Then remove your hand. The
should now cease (LED goes out, acoustic alarm
stops sounding).
After two successive inspiration phases the
disconnection
If audio response is enabled, the ventilator
announces "Check ventilation system and
settings!".
7. Switch off MEDUMAT Easy CPR again.
Drop in O
pressure (<2.7 bar O
2
1. Open the valve of the oxygen cylinder
2. Switch on the MEDUMAT Easy CPR.
3. Close the valve on the oxygen cylinder. When
the oxygen pressure in the system has fallen be-
low 2.7 bar, the
triggered.
If audio response is enabled, the ventilator an-
nounces "Check pressure hose system and gas
supply!".
4. Switch off MEDUMAT Easy CPR again.
stenosis
alarm should be triggered.
<2.7 bar O
alarm should be
2
.
stenosis
alarm
)
2
slowly
.