7. Functional check
7.1 Preparation for functional check
MEDUMAT Easy CPR must not be used if the func-
tional checks reveal defects or deviations from the
specified parameters.
First try to correct the error with the help of the in-
formation provided in section "8. Troubleshooting"
on page 59. If this is not possible, have the unit re-
paired by the manufacturer
– WEINMANN Emergency – or by specialists explicitly
authorized by WEINMANN Emergency.
For the functional check, you require:
–
patient's hose system
–
test bag
–
oxygen cylinder
–
soap-and-water solution made from non-
perfumed soap
–
adapter from test set WM 15357
We recommend that you hold reserve stocks of the
following items:
•
washers for the connections;
•
lip membranes for the patient valve.
1. Connect the device to the oxygen cylinder.
2. Connect the patient's hose system to the device.
Functional check
EN
47