Idiomas disponibles

Audio response
"Ventilation pressure limit
45 mbar"
"Ventilation pressure limit
20 mbar"
"Check airways and settings!"
"Device malfunction!"
"Administer alternative
ventilation!"
"Check pressure hose system
and gas supply!"
"Rule out respiratory arrest and
check mask fit."
"Close oxygen cylinder."
"Check ventilation system and
settings!"
"Selected language: English"
"Audio response is off!"
"CPR mode activated! Ventila-
tion triggered manually!"
"CPR mode deactivated!"
"Provide 2 ventilations now!"
"Perform chest compressions
now!"
EN
38
Using the ventilator
Tube ventilation mode is selected.
Maximum ventilation pressure for tube ventilation.
Mask ventilation mode is selected.
Maximum ventilation for mask ventilation.
MEDUMAT Easy CPR has measured excessive airway resistance.
Check the airways or adjust respiratory frequency and tidal volume
settings to suit the patient (page 24).
The device is faulty or the battery is failing. The device can no longer
be used for ventilation. You must therefore use another ventilation
method (page 41).
MEDUMAT Easy CPR has measured low pressure on the inlet side.
Check whether the O
that the oxygen hose is not leaking, kinked or jammed.
In Demandflow mode:
MEDUMAT Easy CPR can no longer detect a breathing pulse (trig-
ger). Check the patient's breathing, and if necessary switch to a dif-
ferent ventilation mode. Check the connections and mask fit.
In CPR mode without the metronome:
No respiratory stroke has been triggered for 45 seconds. Trigger at
least one respiratory stroke by pressing the key on the MEDUtrigger.
After switching off the ventilator, turn off the O
nal O
supply.
2
Disconnection: a pressure rise of 8 mbar is not achieved during the
inspiration phase under controlled ventilation. This is usually due to an
interruption of the ventilation system or to a low tidal volume setting.
Check the connections or adjust the tidal volume to suit the patient.
When selecting the language for the audio response, press the
mask/ tube switch as many times as necessary until the desired lan-
guage is announced.
Confirmation that audio response is deactivated.
Automatic ventilation is stopped. Trigger the respiratory strokes with
the MEDUtrigger at the right time.
MEDUMAT Easy CPR ventilates with the selected frequency.
Trigger 2 respiratory strokes with the MEDUtrigger.
Undertake 30 chest compressions to the beat of the metronome.
Meaning
cylinder still contains sufficient oxygen and
2
cylinder or the exter-
2