5.3 Performing ventilation

To see the relationship between the values, see
diagram "11.4 Relationship between ventilation
parameters" on page 72.
Maximum ventilation pressure
1. Use the mask/tube switch to set the ventilation
pressure. The LED lights up in active mode.
Recommended maximum ventilation pressure:
Mask ventilation
AT
CP R
Audio response enabled:
"Ventilation pressure limit
If the maximum ventilation pressure is reached, e.g.
in cases where compliance is inadequate,
MEDUMAT Easy CPR sets off a stenosis alarm (see
"Stenosis alarm" on page 32).
Note:
The mask/tube switch can only be operated when
the ventilator is switched on.
Ventilation mask
1. Attach the mask to the patient valve.
2. If necessary, insert a Guedel tube to keep the
airway open before putting on the mask.
3. Place the mask over the patient's mouth and
nose.
4. Tilt the head back and use the E-C technique to en-
sure the mask is hermetically sealed.
20 mbar
20 mbar"
Using the ventilator
Tube ventilation
45 mbar
Audio response enabled:
"Ventilation pressure limit
45 mbar"
EN