2.4 Demandflow function
Demandflow setting
10
600
500
0
10
300
12
15
20
150
25
100
65
Freq.(min
)
-1
V
(ml)
t
2.5 Patient hose system with patient valve
Respiration hose connection
Expiration
tube
Connection
for pressure
gauge tube
Mask/tube connection
EN
12
Description of device
fitted with a lip membrane that enables expired gas
to be conducted away through the expiration tube.
You can check the course of ventilation at ventilation
pressure gauge.
The Demandflow setting switches the
MEDUMAT Easy CPR breathing-controlled O
lation. Such inhalation must be carried out with the
800
respiration mask. A small inspiration (trigger-) pulse
950
causes oxygen to continue flowing until slight over-
pressure interrupts the flow. Expiration then takes
place via the patient valve as in ventilation.
The respiratory gas is routed to the patient via the
patient hose system with patient valve. The patient
hose system with patient valve is designed so that
even if the ventilator fails, spontaneous breathing is
possible, regardless of which ventilation mode you
have selected.
Sponta-
neous
breath-
ing tube
inha-