Alternative Ventilation Procedures - Weinmann MEDUMAT Easy CPR Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para MEDUMAT Easy CPR:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Minute volume (MV) = respiratory frequency x tidal volume (V
For the above example:
Minute volume (MV) = 10 min
Real Demandflow time (min) =
Example:
O
content = 2000 l, MV = 15 l/min.
2
This gives the following equation:
Real Demandflow time =

5.14 Alternative ventilation procedures

-1
x 1.5 l = 15 l/min
oxygen content (l)
MV (l/min)
2,000 l
15 l/min
If MEDUMAT Easy CPR ceases to function during a
ventilation procedure, the following alternatives can
be applied:
Ventilation bags
1. Remove the patient valve from the tube or the
mask.
2. Replace it with the ventilation bag, e.g. a
WEINMANN Emergency COMBIBAG
WM 11000, and perform manual ventilation.
Exhaustion of oxygen supply
In emergency situations when the oxygen supply
runs out, MEDUMAT Easy CPR can also be operated
with respiratory air.
)
t
= 133 min = 2 h 13 min
Using the ventilator
EN