1 Warning and safety information
Information on avoiding, recognizing and elimi-
nating unintended electromagnetic effects
Interference of dental medical devices caused by
radio telephones
Unit with built-in video camera and monitor
Quality of water / air supply
Water quality compliance
HF surgery
Identification of warning and safety information
Smbols used
Ventilation slots
8
+
C1
is a class A device (classified according to CISPR
11, EN 60601-1-2: 1993 based on IEC 60601-1-2). This
device may be operated in a residential area provided it
is used under the responsibility of a medical specialist.
To guarantee the operational safety of electromedical
devices, the use of mobile radio telephones must be
prohibited in the practice or hospital area.
Electrical disturbances in the environment of the C1
(e.g. lightning) may result in reduced image quality. In
such cases, it is advisable to repeat the exposure before
saving the image.
Image artifacts may possibly occur in the event of
radio-frequency interference caused by the transmitters
of e.g. radio services or radio amateurs.
Air and water supply must meet the requirements speci-
fied in the Installation Instructions. Use only clean
water.
To ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. As owner of the treatment center,
you are responsible for the water quality and may have
to take alternative measures to ensure its compliance if
you operate the treatment center without a disinfection
system.
Please contact your specialized dealer or your relevant
dental association for the respective national require-
ments and measures.
This dental treatment center is available both with and
without a high-frequency surgical device (optional).
Only in the Federal Republic of Germany The user is
obligated to keep a "medical product log" if a HF
surgical device is installed! ( see 22.3 "Safety tests")
To avoid personal injury and material damage, you must
also observe the warning and safety information pro-
vided in the present Operating Instructions. They are
i
specially identified with
WARNING.
Observe the accompanying papers (on the chair
nameplate)
Under no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be
obstructed.
Do not spray disinfectants or other similar products into
the ventilation slots.
Sirona Dental Systems GmbH
+
NOTE,
CAUTION or
59 06 446 D 3386
D 3386.201.01.08.02