Sirona Dental Systems GmbH
Chair and dentist element
Dentist and assistant element
Maximum load
Vacuum system
Patient chair
Maintenance of the treatment center
59 06 446 D 3386
D 3386.201.01.08.02
1 Warning and safety information
Please make sure that no foreign parts (e.g. tray) project
into the movement range of the chair and dentist ele-
ment. This is necessary to avoid collisions during auto-
matic movement sequences!
In order to avoid damage to the dentist and assistant ele-
ments and to the chair upholstery, make sure that these
elements do not protrude into the area through which the
chair moves. The ventilation slots must be free!
The maximum load on the dentist and assistant element
may be 1 to 3kg. If the load is more than 1kg, the spring
force must be readjusted by the service engineer (see
Installation Instructions).
The suction of aluminum and other metal oxides from
blasting devices via the automatic separator integrated
in the treatment center and the amalgam separator is
prohibited!
This would cause extreme wear and clogging of the suc-
tion and water paths.
A separate vacuum system must be used in connection
with metal oxide blasting devices.
Treatment centers equipped with central wet aspiration
are generally suitable for suction removal of the above
material. However, please be sure to observe the
instructions provided by the manufacturer of your vac-
uum system.
No restrictions apply when using salt blasting devices in
connection with Sirona treatment centers. However, in
such cases, make sure that the system is subsequently
rinsed with an adequate amount of water.
Please observe the maximum load capacity of the chair
of 135kg according to EN ISO 6875 (tested with four
times safety)
Despite the outstanding quality of your treatment center
and regular maintenance and care by the practice per-
sonnel, it is required in the interest of operational safety
that preventive maintenance is performed at predeter-
mined intervals.
In order to guarantee the operational safety and
functional reliability of your treatment center and to
avoid damage due to natural wear, as the system
owner it is necessary that you have your dental
depot dispatch an authorized service engineer to
check your unit according to the appropriate main-
tenance certificate at regular intervals (at least once
a year).
Please contact your dental depot to obtain a mainte-
nance offer.
9