Device Test - Evaluation; Evaluation Of Device Test Results - Laerdal Serres Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31

Device Test - Evaluation

Evaluation of Device Test Results

After the test is completed, the Vacuum Indicator
will display the results. Press the Test Button to scroll
through the results of each test to display the results.
Test No.
Test 1 -
Occlusions
Test 2 – Vacuum
efficacy
Test 3 – Maxi-
mum vacuum
Test 4 - Leakages
After evaluating the test results turn the Operating Knob to "0" to exit the Device Test.
Test result indication
Test Passed
<100 mmHg
Test Passed
>300 mmHg
Test Passed
>500 mmHg
Test Passed
>450 mmHg
Action if test failed
Check possible blockages (e.g. twisted tub-
ing, blocked filter, blocked filter in the liner)
and run the Device. Test again.
If the High Efficiency Filtration Kit is installed
the pass limit is 150mmHg.
Check Connectors, Tubes and Canister Lid
for leakage* or damage.
Check exhaust outlet for occlusion and run
the Device Test again.
Check Connectors, Tubes and Canister Lid
for leakage* or damage.
Check exhaust outlet for occlusion and run
the Device Test again.
Check Connectors, Tubes and Canister Lid for
leakage* or damage and run the Device Test
again.
16

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

78 00 30

Tabla de contenido