Specifications - Laerdal Serres Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31
Classification
Electrically powered medical suction equipment for field and transport use, according to
ISO10079-1.
High vacuum/high flow.
The LSU is designed for use in road ambulances in accordance with IEC 60601-1-12.
Not suitable for use in the presence of flammable liquids or gases.
Internally powered/class ll equipment type BF, according to IEC 60601-1
The degree of protection provided by the chassis is according to IP34:
Protected against solid foreign objects of 2.5 mm Ø and greater.
Protected against splashing water.
Protected against access with a wire.
General tolerance
Overall tolerance ±5%
Dimensions
Size (h x w x d)
Weight
Canister Capacity
Canister graduation accuracy
Patient Suction Tubing (non-sterile) Cat.No 770410: 8 mm (0.315 in.)
inside diameter x 1.5 m (59 in.) length.
Temperature and Environment
Operating/Charging Temperature
Recommended Charging Temperature
Long term Storage Temperature
Max. 24 hour Storage Temperature
The time required for the LSU to warm from the minimum storage temperature between uses
until it is ready for intended use is min 90 minutes at room temperature.
The time required for the LSU to cool from the maximum storage temperature between uses
until it is ready for intended use is min 90 minutes at room temperature.
Humidity (Operating & Storage)
Altitude
Battery and Charging
Operating/charging AC
Operating/charging DC
315 mm x 330 mm x 160 mm (12.4 in x 13 in x 6.3 in)
4 kg. (8.4 lbs) (including battery NiMH)
1000 ml
±5% of full scale
0 °C to 40 °C (32 °F to 104 °F)
15 °C to 25 °C (59 °F to 77 °F)
0 °C to 40 °C (32 °F to 104 °F)
-30 °C to 70 °C (-22 °F to 158 °F)
5 - 95% RH non-condensing
0 - 4000 m
** 100-240 VAC, 50-60 Hz
** 12-28 VDC
23

Specifications

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

78 00 30

Tabla de contenido