Electromagnetic Conformity - Laerdal Serres Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31

Electromagnetic Conformity

Laerdal Suction Unit is intended for use in the following environments: Professional
Healthcare Facility environment and Emergency Medical Services environment.
Essential performance of the LSU is identified as connection of the patient hose to the
exhaust outlet. This is prevented by coding of the outlet and identifying the exhaust using
a label on the device. EMC disturbances cannot affect this behavior.
No particular actions are required to maintain safety and performance with regard to
electromagnetic disturbances for the expected service life.
Warnings
• Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
• Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the LSU, including cables specified by the Laerdal Medical. Otherwise, degradation of
the performance of this equipment could result.
27
Specifications

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

78 00 30

Tabla de contenido