Troubleshooting for Leakages
If the device test has failed, check whether the system is leaking. Run the Device Test
again blocking different parts in turn, until you find the failure.
1
2
Note
If the LSU does not pass one or more of the steps in this test after suggested actions are
taken, the device might need to be returned for service (see the Troubleshooting guide).
Test by blocking the Pump System
Run the Device Test whilst blocking the outlet. If
the device passes the test, there are no leakages in
the Pump System.
Test by blocking the Patient Tubing Inlet
Run the Device Test whilst blocking the Patient Tubing
inlet on the Canister. If the device passes the test,
there are no leakages in the Canister.
17