Specifications
Electromagnetic Emissions Tests
Emissions Test
RF emissions
Harmonic emissions
Voltage fluctuations/
flicker emissions
Electromagnetic Immunity Tests
Immunity Test
Electrostatic discharge
Radiated RF EM fields
Proximity fields from RF
wireless communications
equipment
Rated power frequency
magnetic fields
Electrical fast transients /
bursts, AC power port
Surges: Line-to-line, AC
power port
Conducted disturbances
induced by RF fields, AC
power port
Voltage dips, AC power
port
Voltage interruptions, AC
power port
Electrical transient con-
duction along supply lines,
DC power port
Standard or test method
CISPR 11
IEC 61000-3-2
IEC 61000-3-3
Standard or test method
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
ISO 7637-2
28
Compliance
Group 1 Class B
Class A
Complies
Compliance Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
380-390 MHz: 27 V/m
430-470 MHz: 28 V/m
704-787 MHz: 9 V/m
800-960 MHz: 28 V/m
1700-1990 MHz: 28 V/m
2400-2470 MHz: 28 V/m
5100-5800 MHz: 9 V/m
30 A/m
50 Hz or 60 Hz
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ± 1 kV
3 V; 0.15 MHz – 80 MHz
6 V in ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT;
25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
Test pulse severity level: III in
Table A2 of ISO 7637-2