Integra Codman ICP EXPRESS Manual De Instrucciones página 15

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 61
Guidance and Manufacturer's Declaration Regarding Electromagnetic Immunity – Non-life Supporting
The Model 82-6634 ICP EXPRESS Monitor is intended for use in the electromagnetic environment specified below. The customer or user of the Model 82-6634 ICP
EXPRESS Monitor must ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms from
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Reference Table 204 IEC 60601-1-2
For RF Immunity Wireless Communication Equipment refer to test method per Table 9 in IEC60601-1-2 4th ed.
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Model 82-6634 ICP EXPRESS Monitor
The Model 82-6634 ICP EXPRESS Monitor is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or
user of the Model 82-6634 ICP EXPRESS Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
Communications Equipment and the Model 82-6634 ICP EXPRESS Monitor as recommended below, according to the maximum output power of the communications
equipment.
Maximum
Output Power
(Watts)
0.01
0.1
1
10
100
Reference Table 206 IEC 60601-1-2
Compliance Level
V1 = 3 Vrms
E1 = 3 V/m
(Equipment and Systems that are NOT Life-Supporting)
Recommended Separation Distances for the Model 82-6634 ICP EXPRESS Monitor
150 kHz to 80 MHz
d = 1.1667
0.11667
0.36894
1.1667
3.6894
11.667
Where P is the maximum output power rating in watts and d is the recommended separation distance in meters.
Table 3
Electromagnetic Environment Guidance
Portable and mobile communications equipment must be separated
from the Model 82-6634 ICP EXPRESS Monitor by no less than the
recommended separation distances calculated/listed below:
d = [3.5/V1]
d = [3.5/E1]
80–800 MHz
d = [7/E1]
800 MHz – 2.5 GHz
Where P is the maximum output power rating in watts and d is the
recommended separation distance in meters.
Field strength from fixed transmitters as determined
by an electromagnetic site survey, must be less than the compliance
levels (V1 and E1).
Interference can occur in the vicinity of equipment containing a
transmitter.
Table 4
80 MHz to 800 MHz
d = 1.1667
0.11667
0.36894
1.1667
3.6894
11.667
10
EN – ENGLISH
800 MHz to 2.5 GHz
d = 2.3333
0.23333
0.73785
2.3333
7.3785
23.333

Publicidad

Tabla de contenido
loading

Tabla de contenido