P1.6 AUTO CALIB. C. - Time of automatic adjustment
Determination of time interval for innitiation of automatic adjustment process (function
unavailable for verified balances).

7.3. MANUAL ADJUSTMENT

7.3.1. Internal Adjustment

The balance performs internal adjustment automatically. During adjustment process do not load the
weighing pan with any load. The message <DO NOT TURN OFF CALIBRATION> is displayed in the
bottom line. Once adjustment process is completed, balance saves its result in memory and returns
to weighing mode.
CAUTION!
- Press Esc button to abort adjustment process.
- If a weighing pan is loaded, message informing about an error turns out on the display. Adjustment
process is aborted, balance returns to the weighing mode. Adjustment process may be repeated upon removal
of the excess load.

7.3.2. External Adjustment

The external adjustment for PS series balances should be carried out with an external mass standard
of class F
(function unavailable for verified balances).
1
Procedure:
• Run an external adjustment process, the balance displays a command ordering to unload the
weighing pan, <REMOVE MASS> (the weighing pan must be empty). When the weighing pan
is unloaded, press
• The balance determines mass of an empty pan, message <CALIBRATION> is displayed in the
bottom line. Next, message <PLACE MASS> and mass value to be placed on the weighing
pan are displayed, e.g. 200.000g (depending on the type of balance).
• Place an external adjustment weight of displayed mass value and press
balance determines the mass, message <CALIBRATION> is displayed in the bottom line. On
completing adjustment process the balance returns to submenu P1.2 EXT.CALIB.

7.3.3. User Adjustment

The external adjustment for PS series balances should be carried out with an external mass standard
of class F
(function unavailable for verified balances).
1
Procedure:
• Run an external adjustment process, the first step of the process is to declare the mass of a
weight that is to be used for adjustment. The mass must be ≥ 30% Max capacity.
• Once the mass of the weight is entered and confirmed, the message prompting the user to
remove the weight from the pan is displayed: <REMOVE MASS> (the weighing pan must be
empty). Unload the pan and press
• The balance determines the weight of an unloaded pan, message <CALIBRATION> is shown
in the bottom line. Next, message <PLACE MASS> and mass value to be placed on the
weighing pan are displayed, e.g. 200.000g (depending on the type of balance).
• Place an external adjustment weight of displayed mass value and press
balance determines the mass, message <CALIBRATION> is displayed in the bottom line. On
completing adjustment process the balance returns to submenu P1.2 EXT.CALIB.