Means Of Operation - RADWAG R Serie Manual Del Usuario

Ocultar thumbs Ver también para R Serie:
Tabla de contenido

Publicidad

F Shortcut Keys
You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.
For procedure informing you how to declare the particular function, read F Shortcut Keys section.

11.13.2. Means of Operation

Enter <PIPETTES CALIBRATION.> mode.
Start the process.
Follow the prompts to set the correct options.
Enter the ambient temperature, next press ENTER for confirmation. The computer software
proceeds to the next step, i.e. ambient humidity settings.
Set the ambient humidity, next press ENTER for confirmation. The computer software proceeds
to the next step, i.e. air pressure settings.
Set the air pressure, next press ENTER for confirmation. The computer software proceeds to
the next step, i.e. entering the first control volume (V1) for a tested pipette.
Enter (V1) control volume. For fixed volume pipettes (P2.13.5 VOLUME DETERMINATION
parameter set to value <1>) this is the one and only value to be entered. The computer
software proceeds to the next step, i.e. entering the second control volume (V2) for a tested
pipette.
Enter (V2) control volume. The computer software proceeds to the next step, i.e. entering the
third control volume (V3) for a tested pipette.
Enter (V3) control volume.
- 68 -

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

As r2Ps r1Ps r2Ps r2.hApp r

Tabla de contenido