Send stable measurement result in a basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2-3
4
stability
S
space
marker
An example:
S CR LF
- command send form a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR
where: _ - space
Immediately send measurement result in a basic measuring unit
Format: SI CR LF
Response options:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in a basic measuring unit
Response format:
1
2
3
4
S I
space
An example:
S I CR LF
– command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
where: _ - space
Send stable measurement result in a current measuring unit
Format: SU CR LF
Response options:
SU_A CR LF
- command understood and in progress
SU_E CR LF
- time limit exceeded while waiting for a stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
S
U
space
An example:
S U CR LF
S U _ A CR LF
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
where: _ - space
5
space
5
stability
space
marker
4
5
stability
space
marker
- command send form a computer
- command understood and in progress
6
7-15
character mass
- command carried out, response: mass value in a basic
LF
measuring unit.
6
7-15
character mass
- command carried out, immediate response: mass
value in a basic measuring unit
6
7-15
character
mass
- command carried out, response: mass value in a
current measuring unit
- 85 -
16
17
18
space
unit
16
17
18
space
unit
16
17
18
19
space
unit
19
20
21
CR
LF
19
20
21
CR LF
20
21
CR
LF