RADWAG R Serie Manual Del Usuario página 90

Ocultar thumbs Ver también para R Serie:
Tabla de contenido

Publicidad

working mode.
Format: UI CR LF
Response options:
UI_"x
,x
, ... x
"_OK CR LF
1
2
n
UI_I CR LF
x – unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
Response:
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x CR LF
Response options:
US_ x_OK CR LF
US_E CR LF
US_I CR LF
x – parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „
pressing).
An example:
Command:
Response:
Give current unit
Command overview: command returns current unit.
Format: UG CR LF
Response options:
UG_x_OK CR LF
UG_I CR LF
x – parameter, unit symbol
An example:
Command:
Response:
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS CR LF
Response options:
BP_OK CR LF
BP_E" CR LF
BP_I CR LF
TIME – parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350 CR LF
Response:
BP_OK CR LF
CAUTION!
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.
UI CR LF
UI_"g, mg, ct"_OK CR LF
– command carried out, response: current unit
– error in-course of command execution, no parameter or incorrect
format
– command understood but not accessible at this moment
US_mg CR LF
US_mg_OK CR LF
– command carried out, response: current unit
– command understood but not accessible at this moment
UG CR LF
UG_ct_OK CR LF
– command carried out, BEEP sound signal activated
– no parameter or incorrect format
– command understood but not accessible at this moment
– command carried out, returns accessible units
– command understood but not accessible at this
moment
– return available units
– response: available units
– set „mg" unit
– „mg" set as a current unit
– return current unit
– currently set unit is „ct"
– activate BEEP for 350 ms
– BEEP activated
- 90 -