Precision Novi™ Implantable Pulse Generator
Specifications and Technical Data
Parameter
Areas (Channels)
Amplitude
Rate
Width
Cycle
Ramp ON
Contacts
a. Only one Area is available if the rate is >130 pps.
b. Amplitude × Width ≤ 12.7 µC.
Materials
Case
Header
Strain Relief
Size/Volume
Radiopaque Identification Tag
The IPG contains a radiopaque identification tag "BSC S1140". The identification tag is visible
using standard x-ray procedures.
Registration Information
In accordance with international practice and regulatory legislation in some countries, a
registration form is packed with each Boston Scientific neurostimulator. The purpose of this
form is to maintain traceability of all products and to secure warranty rights. It also allows the
institution involved in the evaluation or replacement of a specific device to gain quick access to
pertinent data from the manufacturer.
Boston Scientific Neuromodulation
25155 Rye Canyon Loop
Valencia, California 91355, USA
Attention: Customer Service Department
Precision Novi™ Implantable Pulse Generator Directions for Use
90930910-02 REV B
4 of 239
Range
4
0 – 25.5 mA
2 – 1200 pps
a
20 – 1000 μsec
b
1 sec – 90 min, OFF
1 – 10 sec
1 – 16, case: +100% to -100%, OFF
Titanium
Epoxy
Silicone
70.9 mm x 49.5 mm x 11.3 mm / 33 cm
Default
—
0 mA
40 pps
210 μsec
OFF
3 sec
1 – 16, case: OFF
(including header)