Information for Prescribers
Recommended separation distances between portable and mobile RF
communications equipment and the Precision Montage MRI System
The Precision Montage MRI System is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Precision Montage MRI
System can help prevent electromagnetic interference by maintaining a minimum distance of 30 cm
between portable and mobile RF communications equipment (transmitters) and the Precision Montage
MRI System.
NOTE These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Notice to the User Per Industry Canada Radio Standards Specifications:
This device complies with Industry Canada license-exempt RSS standard(s). Operation is
subject to the following two conditions:
1. this device may not cause interference, and
2. this device must accept any interference, including interference that may cause undesired
operation of the device.
Technical Service
There are no user serviceable parts. If you have a specific question or issue, please contact
your Boston Scientific representative.
End of Programmed Service
The Precision Montage MRI System IPG software is programmed to end service after 12
years. As the IPG nears the end of the programmed period, the Remote Control and Clinician
Programmer provide the following indicators to inform the user that the end of the programmed
period is approaching:
• Remote Control - Approximately six months before the end of programmed
period, the Remote Control displays a weekly message indicating the
number of service days remaining. Approximately one month before
the end of the programmed period, the message displays daily.
Precision Montage™ MRI System Information for Prescribers