Insufflator
SIDNE Says
Insufflator...
High flow
Medium flow
Low flow
Activated
Stopped
Pressure set to 15
Overpressure
Leakage
Occlusion
Low gas level
Gas supply empty
Heater error
Requires service
Meaning
Added as a prefix to insufflator-related messages.
The insufflator has been set to high flow.
The insufflator has been set to medium flow.
The insufflator has been set to low flow.
The insufflator has been activated.
The insufflator has been stopped.
The insufflator pressure has been set to 15 mmHg.
The insufflator actual pressure is 5 mmHg higher than
the selected pressure, or the actual pressure is equal to or
higher than 30 mmHg.
This error message is played 15 seconds after the usual
insufflator error sound occurs.
The insufflator senses a leak in the pneumoperitoneum.
This error message is played 15 seconds after the usual
insufflator error sound occurs.
The insufflator senses a block in the tube, Veress
cannula, or trocar.
This error message is played 15 seconds after the usual
insufflator error sound occurs.
The pressure of the gas-supply cylinder is low (<25 bar).
The gas-supply cylinder is empty.
The gas temperature is higher than 41°C.
The insufflator requires service from an
authorized technician. Contact technical support
(+1.800.624.4422).