3.9
Danger of burns
Argon plasma can strongly heat electrically conductive materials,
e.g. metal. To avoid burns, never direct argon plasma towards
electrically conductive materials.
3.10 Risk of accidental tissue damage with the APC 300
Model
If the gas tanks of the APC 300 Model are not opened or are
completely empty, a spray coagulation without argon plasma is
performed with the instrument after each activation. There is an
audible warning signal; the tanks on the display of the APC 300
Model blink and an error message appears. In the spray coagu-
lation mode there is a danger of unintentional tissue damage;
the instrument can also be damaged!
Before using the APC 300 Model always open the gas tanks. Ob-
serve the self-check performed by the Unit. How much gas is
there left in the tanks? The tank symbols on the display of the
APC 300 Model give an approximate indication of the quantity of
gas remaining.
Never work with two gas tanks which are almost completely
empty. When one gas tank is empty, the APC 300 Model switches
over to the second gas tank automatically once the regulator is
connected.
Always observe the warning signals and error messages.
3.11 Modification of the product by the user
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.