OBJ_DOKU-189911-006.fm Seite 32 Donnerstag, 17. Juni 2021 9:31 09
8.6
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Completely fill a disposable syringe with cleaning solution
and rinse the lumina (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
8. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
9. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
10. Rinse the surfaces with sterile, deionized water (at least
1 minute).
32