Descargar Imprimir esta página

Required Aids; Precleaning - Erbe 20132-106 Manual Del Usuario

Apc-aplicadores

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 72
OBJ_DOKU-189911-006.fm Seite 31 Donnerstag, 17. Juni 2021 9:31 09
8.4

Required aids

Processing step
Items Needed

Precleaning

Soft plastic brush/soft cloth
10 ml disposable syringe with Luer-Lock
connection
Erbe irrigation adapter 20132-095
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
10 ml disposable syringe with Luer-Lock
connection (4 ea.)
Erbe irrigation adapter 20132-095
Compressed air gun (2 bar maximum)
Machine cleaning/
Erbe irrigation adapter 20132-095
disinfection
8.5
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Screw the irrigation adapter (Art. No. 20132-095) onto the
APC applicator.
3. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
4. Remove the syringe. Leave the irrigation adapter on the
product for the next processing steps.
31

Publicidad

loading

Este manual también es adecuado para:

20132-110