Erbe Elektromedizin expressly warns against modifying the ca-
ble, e.g. independently replacing defective insulation. Any mod-
ification leads to the termination of liability on behalf of Erbe
Elektromedizin.
4

How to use

Clean, disinfect and sterilize these products before using them
for the first time and before every subsequent use.
WARNING! Do not kink the cable.
WARNING! Protect the product from all physical damage! Do not
throw! Do not apply force of any kind!
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the cable.
Connect the instrument to the cable. Then plug the connector on
the cable into a suitable socket on the electrosurgical unit.
Select an appropriate unit setting on the electrosurgical unit. The
unit setting must not exceed the maximum electrical capacity of
both the cable and the instrument as this could damage the ac-
cessories. Information on selecting appropriate unit settings is
provided in the User Manual of the electrosurgical unit.
5

Cleaning, disinfection, sterilization

5.1

Safety instructions for processing

May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
23